Flow system – RKI Instruments 35-3010RK-03 User Manual

Page 7

Advertising
background image

35-3010RK-03 Sample-Draw Detector • 3

Flow System

The sample draw detector’s flow system consists of the INLET fitting, hydrophobic filter,
charcoal filter, pump, flowmeter, bypass valve, status lights, pressure switch, flow block,
and EXHAUST fitting (see Figure 1). Figure 2 illustrates how the gas sample moves
through the flow system. The sample flow to the sensors is controlled by a bypass valve so
that the total flow is more than that indicated by the flowmeter. This reduces the time for
the sample to reach the sample draw detector from the sampling area.

Figure 2: Sample Draw Detector Flow Diagram

INLET fitting

The INLET fitting on the bottom of the housing allows the gas sample to enter the sample
draw detector. The INLET fitting accepts 1/4 in. rigid tubing. See “Installation” on page 8
for instructions to connect tubing to the INLET fitting.

Hydrophobic Filter

The hydrophobic filter is to the left of the main circuit board. It is held in place by a metal
clip. It prevents water and other liquids from contaminating the flow system. Replace the
filter when it appears dirty, discolored, or clogged. If a liquid other than water is drawn
into the filter, replace the filter as soon as possible.

Hydrophobic

Filter

Inlet

Exhaust

Pump

Pressure Switch

Oxygen

Flow meter

H2S

Flow Block

Bypass

Valve

CO

Charcoal

Filter

LEL

Advertising