RKI Instruments 35-3010RK-04 User Manual

Page 18

Advertising
background image

14 • 35-3010RK-04 Sample-Draw Detector

1.

Verify that the controller is reading 0 for the combustible and CO channels.

If the reading is not 0 on any of the channels, set the zero reading as described in
“Start Up” on page 12, then continue this procedure.

2.

Assemble the calibration kit as described in the Calibration section of this manual.
Use of a 3-gas cylinder is recommended so that both channels may be checked at once.

Performing the response test

NOTE:

This procedure describes the RKI calibration kit that includes a demand flow
regulator.

1.

Screw the regulator into the calibration cylinder.

2.

Connect the calibration tubing from the regulator to the inlet line at or near the INLET
fitting. Gas will begin to flow.

3.

After approximately one minute, verify that the reading for each channel at the
controller stabilizes within ± 10% of the concentration of the test sample. If the
reading is not within ± 10% of the test sample, calibrate the sample draw detector as
described in the Calibration section of this manual.

4.

Remove the calibration tubing from the inlet line, then reconnect the inlet line.

5.

Store the calibration kit in a safe place.

Quarterly

Calibrate the sample draw detector as described in the “Calibration” section.

Troubleshooting

The troubleshooting guide describes symptoms, probable causes, and recommended
action for problems you may encounter with the sample draw detector.

NOTE:

This troubleshooting guide describes sample draw detector problems only. See
the instruction manual for the controller if it exhibits any problems.

Fail Condition

Symptoms

The sample draw detector’s Fail light is on.

The monitoring device is operating properly but indicates a reading well below zero
on one or more channels.

Probable Causes

The sample draw detector’s flow rate is too low because of an obstructed sample line,
failed pump, etc.

The sample draw detector is malfunctioning.

The sensor or transmitter wiring is disconnected or misconnected.

Recommended Action

1.

At the sample draw detector, set the correct flow rate with the bypass valve or flow
adjust potentiometer.

2.

If you cannot set the correct flow rate, check the sample lines for obstructions or kinks.

3.

Verify that the sensor wiring is correct and secure. “Wiring the Sample Draw

Advertising