18 • 35-3010RK-04 Sample-Draw Detector

5.

When the instructions call for applying gas to the detector, connect the sample tubing
from the regulator to the inlet line at or near the INLET fitting.

6.

Allow the gas to flow for 1 minute.

7.

Set the zero reading according to the controller operator’s manual.

8.

Disconnect the sample tubing from the inlet line.

9.

Unscrew the regulator from the zero air calibration cylinder. Leave the sample tubing
connected to the regulator.

Setting the Response Reading

1.

Screw the regulator into the 4-gas calibration cylinder.

2.

Connect the sample tubing to the regulator.

3.

Follow the directions in the controller’s operator’s manual for setting the response
reading for the LEL channel.

4.

When the instructions call for applying gas to the detector, connect the sample tubing
from the regulator to the inlet line at or near the INLET fitting.

5.

Allow the calibration gas to flow for 1 minute.

6.

Set the response reading according to the controller operator’s manual to match the
calibration gas concentration.

7.

Disconnect the sample tubing from the inlet line.

8.

Unscrew the regulator from the calibration cylinder.

NOTE:

For convenience, leave the regulator connected to the sample tubing.

Returning to Normal Operation

1.

Remove the voltmeter leads from the test points.

2.

Reconnect the incoming sample line.

3.

Wait 1 to 2 minutes to allow the calibration gas to be drawn out and the reading to
stabilize.

4.

Close the housing door.

5.

Store the components of the calibration kit in a safe and convenient place.

Calibration, CO Detector

This section describes how to calibrate the CO detector in the sample draw detector. It
includes procedures to prepare for calibration, set the zero reading, set the response
reading, and return to normal operation.

NOTE:

This procedure describes calibration using a demand flow regulator, a zero air
calibration cylinder, and a 3-gas calibration cylinder.

Preparing for Calibration

CAUTION:

This procedure may cause alarms at the controller. Take appropriate action to avoid
this, such as entering the calibration mode at the controller.