Calibration frequency, Calibration, h, S detector – RKI Instruments 35-3010RK-07 User Manual

Page 22

Advertising
background image

18 • 35-3010RK-07 Sample-Draw Detector

Calibration Frequency

Although there is no particular calibration frequency that is correct for all applications, a
calibration frequency of every 3 months is adequate for most sample draw detector
applications. Unless experience in a particular application dictates otherwise, RKI
Instruments, Inc. recommends a calibration frequency of every 3 months for the oxygen
and H

2

S sensors and every 6 months for the IR CH

4

sensor in the sample draw detector.

If an application is not very demanding, for example detection in a clean, temperature
controlled environment, and calibration adjustments are minimal at calibration, then a
calibration frequency of every 6 months for the oxygen and H

2

S sensors and every 12

months for the IR CH

4

sensor is adequate for the sample draw detector.

If the application is very demanding, for example if the environment is not well
controlled, then more frequent calibration than every 3 months for the oxygen and H

2

S

sensors and every 6 months for the IR CH

4

sensor may be necessary for the sample draw

detector.

Calibration, H

2

S Detector

This section describes how to calibrate the H

2

S detector in the sample-draw detector. It

includes procedures to prepare for calibration, set the zero reading, set the response
reading, and return to normal operation.

NOTE:

This procedure describes calibration using a demand flow regulator, a zero air
calibration cylinder, and a 4-gas calibration cylinder.

Preparing for Calibration

CAUTION:

This procedure may cause alarms at the controller. Take appropriate action to avoid
this, such as entering the calibration mode at the controller.

1.

Open the housing door.

2.

Set a voltmeter to measure in the millivolt (mV) range.

3.

Plug the voltmeter leads into the test points in the AMP 2 section of the main circuit
board. Plug the positive lead into the test point labeled CAL+2; plug the negative lead
into the test point labeled CAL-2.

4.

Use the following formula to determine the correct test points output for the H

2

S

calibrating sample.

Output (mV) = (calibrating sample/fullscale) X 400 + 100

For example, with a calibrating sample of 25 ppm H

2

S and a fullscale setting of 100

ppm H

2

S, the correct output for the H

2

S test points is 200 mV.

200 (mV) = (25/100) X 400 +100

Advertising