2 • 35-3010RK-07 Sample-Draw Detector

Description

This section describes the components of the

35-3010RK-07 sample draw detector

. The sample

draw detector consists of the housing, flow system, and detection system.

Figure 1: Sample Draw Detector Component Location

Housing

The sample draw detector’s fiberglass housing is weather- and corrosion-resistant. It is
suitable for installation where general purpose equipment is in use. The housing door is
hinged on the left side and is secured by two latches on the right side.

Four mounting feet are attached to the back of the housing (one at each corner). Use the
mounting feet to install the housing to a vertical surface. Use the two conduit hubs on the
bottom of the housing to make wiring connections.

An aluminum subpanel is mounted to the interior of the housing. The sample draw
detector’s internal components are mounted to the subpanel.

Amp 2

IR CH4
Sensor

Flow Block
for IR CH4

Preamp
Circuit Board

Amp 1

Charcoal Filter

Hydrophobic Filter

(AcroPak)

Flow Block for
Oxygen and H2S

Pump

Oxygen Sensor

H2S Sensor

Dummy Plug for
Combustible Sensor

Exhaust Fitting For 1/4"

OD Rigid Tubing

3/4 Conduit Hub, 2X

Dummy Plug for
CO Sensor

Inlet Fitting For 1/4"

OD Rigid Tubing

Reset Switch

Main Circuit Board