Calibration frequency, Calibration – RKI Instruments 65-2320RK User Manual

Page 15

Advertising
background image

65-2320RK Oxygen Transmitter • 11

8.

Install the detector and controller terminals strips into their sockets on the new
amplifier as shown in Figure 3 on page 6. If controller leads or detector leads were
removed during this procedure, refer to Figure 3 on page 6 for the detector and
amplifier connections.

NOTE:

When a transmitter is first powered up with a new amplifier, the initial output
may be either high or below zero depending on the setting of the zero pot. Be
sure to make arrangements so that this does not cause unwanted alarms.

9.

Turn on power to the controller.

10. Turn on the controller and place it into normal operation.

11. Allow the transmitter to warm-up for 5 minutes.

12. Calibrate the transmitter as described in “Calibration” on page 11 of this manual.

Calibration Frequency

Although there is no particular calibration frequency that is correct for all applications, a
calibration frequency of every 3 months is adequate for most oxygen transmitter
applications. Unless experience in a particular application dictates otherwise, RKI
Instruments, Inc. recommends a calibration frequency of every 3 months for the oxygen
transmitter.

If an application is not very demanding, for example detection in a clean, temperature
controlled environment, and calibration adjustments are minimal at calibration, then a
calibration frequency of every 6 months is adequate.

If an application is very demanding, for example if the environment is not well controlled,
then more frequent calibration than every 3 months may be necessary.

Calibration

This section describes how to calibrate the oxygen transmitter. It includes procedures to
prepare for calibration, set the fresh air reading, set the zero reading, and return to normal
operation. It describes calibration using a calibration kit that includes a calibration cup,
calibration gas, sample tubing, and a fixed flow regulator with an on/off knob. RKI
Instruments, Inc. recommends using a 0.5 LPM (liters per minute) fixed flow regulator.

Preparing for Calibration

1.

Place the controller into its calibration mode or disable external alarms.

2.

Unscrew and remove the junction box cover.

3.

Set a voltmeter to measure in the millivolt (mV) range.

4.

Plug the voltmeter leads into the test points on the amplifier. Plug the positive lead
into the red + test point; plug the negative lead into the black - test point.

Advertising