65-2320RK Oxygen Transmitter • 1

Overview

This instruction manual describes the 65-2320RK oxygen transmitter. This manual also
describes how to install, start up, configure, maintain, and calibrate the transmitter when
using it with a gas monitoring controller. A parts list at the end of this manual lists
replacement parts and accessories for the oxygen transmitter. See the controller operator’s
manual for information specific to the controller.

Specifications

Table 1 lists specifications for the oxygen transmitter.

WARNING: When using the 65-2320RK, you must follow the instructions and warnings

in this manual to assure proper and safe operation of the 65-2320RK and to
minimize the risk of personal injury. Be sure to maintain and periodically
calibrate the 65-2320RK as described in this manual.

Table 1: Specifications

Target Gas

Oxygen

Sampling Method

Diffusion

Detection Range

0 to 25%

Signal Output

4 to 20 mA

Response Time

90% in 30 seconds

Accuracy

± 0.5% O