65-2320RK Oxygen Transmitter • 13

Returning to Normal Operation

1.

Remove the voltmeter leads from the amplifier test points.

2.

Secure the junction box cover to the junction box.

3.

When the controller display reading rises above the decreasing alarm setpoint, return
the controller to normal operation.

NOTE:

If you do not allow the gas reading to increase above the decreasing alarm point,
then unwanted alarms may occur.

4.

Verify that the controller display reading increases and stabilizes at 20.9% volume.

5.

Store the components of the calibration kit in a safe and convenient place.

Parts List

Table 3 lists replacement parts and accessories for the oxygen transmitter.

Table 3: Parts List

Part Number

Description

06-1248RK-03

Calibration kit sample tubing, 3 foot length

18-0400RK-01

Junction box with rubber spacers

57-1064RK-01

S2 series oxygen amplifier, w/orienting gasket, 0 - 25%

65-2320RK

Oxygen transmitter, non-explosion proof, includes detector

65-2510RK

Oxygen detector, non-explosion proof

71-0173RK

65-2320RK Operator’s Manual (this document)

81-0076RK

Zero air calibration cylinder, 17 liter steel

81-0076RK-01

Zero air calibration cylinder, 34 liter steel

81-0076RK-03

Zero air calibration cylinder, 103 liter steel

81-0078RK

100% nitrogen calibration cylinder, 17 liter steel

81-0078RK-01

100% nitrogen calibration cylinder, 34 liter steel

81-0078RK-03

100% nitrogen calibration cylinder, 103 liter steel

81-F301RK

Calibration kit for oxygen detector head, 103 liter

81-F301RK-LV

Calibration kit for oxygen detector head, 34 liter

81-1050RK

Regulator with gauge and knob, 0.5 LPM, for 17 liter and 34 liter steel calibration cylinders

81-1051RK

Regulator with gauge and knob, 0.5 LPM, for 34AL/58/103 liter calibration cylinders

81-1117RK

Calibration cup