Installation – RKI Instruments 65-2395RK-CH4 User Manual

Page 8

Advertising
background image

4 • 65-2395RK-CH4 Methane Transmitter

amplifier. Use the top 3/4’’ conduit hub to connect wiring from the amplifier to the
controller. Use the cover on the front of the junction box to access the interior of the
junction box. The detector and amplifier are factory installed in the junction box. Three
spacers installed on the back of the junction box control the distance of the junction box
from a mounting surface and ensure that there is enough room to install a calibration cup
on the detector during calibration.

Installation

This section describes procedures to mount the methane transmitter in the monitoring
environment and wire the transmitter to a controller.

Mounting the Methane Transmitter

1.

Select a mounting site that is representative of the monitoring environment. Consider
the following when you select the mounting site.

Select a site where the transmitter is not likely to be bumped or disturbed. Make
sure there is sufficient room to perform start-up, maintenance, and calibration
procedures.

Select a site where the target gas is likely to be found first. For lighter gases,
mount the detector near the ceiling; for heavier gases, mount the detector near the
floor.

Figure 2: Mounting the Methane Transmitter

.38

2.70

.75

Rubber
Spacers,
3X

J-Box

3.65

3/4 NPT

Female

IR CH4 % Volume

Detector

7.10 MAX