Calibration frequency, Calibration – RKI Instruments 65-2427RKSS-05 User Manual

Page 16

Advertising
background image

12 • 65-2427RKSS-05 H

2

S Detector

8.

Reinstall the junction box cover.

9.

Turn on or plug in power to the controller.

10. Turn on the controller and place into normal operation.

CAUTION:

Allow the replacement detector to warm up for 5 minutes before you continue with
the next step.

11. Calibrate the replacement detector as described in the Calibration section of this

manual.

Calibration Frequency

Although there is no particular calibration frequency that is correct for all applications, a
calibration frequency of every 3 months is adequate for most H

2

S detector applications.

Unless experience in a particular application dictates otherwise, RKI Instruments, Inc.
recommends a calibration frequency of every 3 months for the H

2

S detector.

If an application is not very demanding, for example detection in a clean, temperature
controlled environment, and calibration adjustments are minimal at calibration, then a
calibration frequency of every 6 months is adequate.

If an application is very demanding, for example if the environment is not well controlled,
then more frequent calibration than every 3 months may be necessary.

Calibration

This section describes how to calibrate the H

2

S detector. It includes procedures to prepare

for calibration, set the zero reading, set the response reading, and return to normal
operation. It describes calibration using a calibration kit that includes a calibration cup,
calibration gas, sample tubing, and a fixed flow regulator with an on/off knob. RKI
Instruments, Inc. recommends using a 0.5 LPM (liters per minute) fixed flow regulator.

Preparing for Calibration

NOTE: Calibrating the H

2

S detector may cause alarms. Be sure to put the controller into

its calibration program or disable external alarms before calibrating.

WARNING: Do not remove the detector housing cap while the circuits are energized

unless the area is determined to be non-hazardous. Keep the detector
housing cap tightly closed during operation.

1.

Screw the calibration cup onto the bottom of the H

2

S detector.

2.

Screw the regulator into the zero air calibration cylinder.

3.

Use the calibration kit sample tubing to connect the fixed flow regulator to the
calibration cup.

4.

Put the controller into its calibration program.