Troubleshooting, Replacing components of the ppm hexane transmitter – RKI Instruments 65-2461RK User Manual

Page 14

Advertising
background image

10 • 65-2461RK PPM Hexane Transmitter

Troubleshooting

The troubleshooting guide describes symptoms, probable causes, and recommended
action for problems you may encounter with the ppm hexane transmitter.

NOTE:

This troubleshooting guide describes transmitter problems only. See the
controller operator’s manual for problems you may encounter with the
controller.

Replacing Components of the PPM Hexane Transmitter

This section includes procedures to replace the ppm hexane detector and amplifier.

Replacing the PPM Hexane Detector

1.

Turn off power to the controller.

2.

Place the controller’s on/off switch in the OFF position.

3.

Remove the junction box cover.

4.

Remove the detector terminal strip from its socket.

5.

Disconnect the detector leads from the detector terminal strip. Note the position of the
color-coded leads as you remove them.

6.

Unscrew the detector from the junction box.

7.

Guide the detector leads of the replacement detector through the bottom conduit hub
of the junction box, then screw the mounting threads of the detector into the conduit

Table 2:Troubleshooting the ppm Hexane Transmitter

Condition

Symptom(s)

Probable Causes

Recommended Action

Fail
Condition

• Controller indicates a

fail condition.

• The transmitter wiring

is disconnected or
misconnected.

• The transmitter’s zero

reading is low enough
to cause a fail
condition.

• The transmitter is

malfunctioning.

1. Verify that the transmitter wiring is

correct and secure.

2. Calibrate the transmitter.
3. If the fail condition continues, replace

the detector.

4. If the fail condition continues, contact

RKI for further instruction.

Slow or No
Response/
Difficult or
Unable to
Calibrate

• Transmitter responds

slowly or does not
respond to response
test.

• Unable to accurately

set the zero or
response reading
during calibration.

• Transmitter requires

frequent calibration.

Note: Under “normal”
circumstances, the
transmitter requires
calibration once every 6
months.

Some applications may
require a more frequent
calibration schedule.

• The calibration cylinder

is low, out-dated, or
defective.

• The calibration gas

flow rate is too low.

• The transmitter is

malfunctioning.

1. Verify that the calibration cylinder

contains an adequate supply of a
fresh test sample.

2. Verify that the regulator used for

calibration is a 0.5 LPM regulator.

3. If the calibration/response difficulties

continue, replace the detector.

4. If the calibration/response difficulties

continue, contact RKI for further
instruction.

Advertising