Description, Ppm hexane detector – RKI Instruments 65-2461RK User Manual

Page 6

Advertising
background image

2 • 65-2461RK PPM Hexane Transmitter

Description

This section describes the ppm hexane transmitter’s components. It is a 4 to 20 mA type
detector head. It consists of the ppm hexane detector, amplifier, and junction box.

Figure 1: PPM Hexane Transmitter Component Location

PPM Hexane Detector

The ppm hexane detector is a MOS type (metal oxide semiconductor) detector packaged
in a 1/2 inch NPT nipple with a sintered metal flame arrestor on one end allowing
ambient air to diffuse into the detector. The flame arrestor also contains any sparks which
may occur within the detector. The 1/2 inch NPT mounting threads at the top of the
detector allow you to mount it into the bottom conduit hub of the junction box. A
rainshield screws onto the bottom of the detector (flame arrestor end). The rainshield
helps protect the detector from rain and debris in the monitoring environment. Three
color-coded leads extend from the top of the detector. The leads allow you to connect the
detector to the amplifier. The output of the detector is non-linear and is linearized by the

Controller Terminal Strip

Hexane Detector

Mounting Screw
w/ flat and lock washer

Detector Terminal Strip

Span Pot

Zero Pot

Factory Adjust Pot

Load Voltage Pot

Factory Adjust Pot

Factory Use

Jumper Strip

Amplifier

J-Box

1-

2

H

e

x

3-

4

H

2

G

a

s

S

e

le

c

t

1

3

4

2

SP

A

N

ZE

R

O

+ Test Point, Red

Reducing Bushing,

3/4 NPT to 1/2 NPT

Gas Select Jumper,

In Hydrogen Position

Factory Use Jumper Strip

- Test Point, Black

WB

R

D

E

T

E

C

T

O

R

P

W

R

/

S

IG

S