Maintenance – RKI Instruments 65-2484RK User Manual

Page 25

Advertising
background image

65-2484RK Multi Point Detector • 21

Maintenance

This section describes maintenance procedures. It includes preventive maintenance,
troubleshooting, and component replacement procedures.

Preventive Maintenance

This section describes a preventive maintenance schedule to ensure the optimum
performance of the multi-point detector. It includes daily, monthly, and quarterly
procedures.

Daily

At the controller, verify a display reading of:

0 %LEL for combustible gas

20.9% for oxygen

0 ppm for H

2

S, CO, or toxic ESM-01 channel

Investigate significant changes in the display reading.

Monthly

This procedure describes a test to verify that each detector in the multi-point detector
responds properly to its target gas. It describes a test using a fixed flow regulator with an
on/off knob. RKI Instruments, Inc. recommends using a 0.5 LPM (liters per minute) fixed
flow regulator for the LEL, O

2

, H

2

S, or CO channels. See Table 3 on page 33 to determine

the correct calibration cup and regulator for the ESM-01 toxic channel.

WARNING: The ESM-01 detector calibration cup is different than that of the LEL/O

2

/

H

2

S/CO detectors. Ensure that you are using the correct calibration cup for

the detector you are testing. See “Calibration, ESM-01 Toxic Detector” on
page 33.

NOTE:

Performing a response test on the multi-point detector may cause alarms. Be sure
to put the controller into its calibration program or disable external alarms before
performing this test.

Preparing for the response test

1.

Place the controller into its calibration program or disable external alarms.

2.

Verify that the controller display reading for the channel you are testing is:

0 %LEL for combustible gas

20.9% for oxygen

0 ppm for H

2

S, CO, or toxic ESM-01 channel

If the display reading is not the appropriate fresh air reading, set the zero reading of
the detector as described in the Start Up section of this manual, then continue this
procedure.

3.

Screw the calibration cup onto the bottom of the detector you are testing if you are
testing an LEL, O

2

, H

2

S, or CO detector. If you are testing the ESM-01 detector,

unscrew the splashguard before pushing on the calibration cup.

4.

Screw the regulator into the calibration cylinder.

Advertising