Troubleshooting – RKI Instruments 65-2497RK User Manual

Page 12

Advertising
background image

8 • 65-2494RK/65-2497RK Oxygen Detector

Troubleshooting

The troubleshooting guide describes symptoms, probable causes, and recommended
action for problems you may encounter with the oxygen detector.

NOTE:

This troubleshooting guide describes detector problems only. See the controller
operator’s manual for problems you may encounter with the controller.

Table 2: Troubleshooting the Oxygen Detector

Condition

Symptom(s)

Probable Causes

Recommended Action

Fail Condition

• Controller indicates a

fail condition.

• The detector wiring is

disconnected or
misconnected.

• The plug-in sensor is

not properly plugged
into the socket pattern
in the detector housing
body.

• The detector’s zero

reading is low enough
to cause a fail
condition.

• The detector is

malfunctioning.

1. Verify that the detector wiring is

correct and secure.

2. Confirm that the plug-in sensor is

installed properly.

3. Perform a fresh air adjustment. A full

calibration is recommended.

4. If the fail condition continues, replace

the plug-in sensor as described later
in this section.

5. If the fail condition continues, contact

RKI for further instruction.

Slow or No
Response/
Difficult or
Unable to
Calibrate

• Detector responds

slowly or does not
respond to response
test.

• Unable to accurately

set the fresh air or
zero reading during
calibration.

• Detector requires

frequent calibration.

Note: Under “normal”
circumstances, the
transmitter requires
calibration once every 3
months.

Some applications

may require a more
frequent calibration
schedule.

• The calibration cylinder

is low, out-dated, or
defective.

• The calibration gas is

not an appropriate
concentration.

• The membrane on the

detector housing cap is
blocked with dirt or
some other particulate
contamination.

• The detector is

malfunctioning.

1. Verify that the calibration cylinder

contains an adequate supply of a
fresh test sample.

2. Verify that the calibration gas

concentration is appropriate for the
transmitter. Zero emission air (20.9%
oxygen) is normally used for a fresh
air adjustment if the environment is
suspect and 100% nitrogen (0%
oxygen) is normally used for a zero
adjustment.

3. Check the sensor face for dirt. Clean

if necessary.

4. Check the face of the detector

housing cap and remove any
particulate contamination from the
hydrophobic membrane if necessary.
If the membrane appears saturated
with contamination or damaged,
replace the membrane as described
in the next section.

5. If the calibration/response difficulties

continue, replace the plug-in sensor
as described later in this section.

6. If the calibration/response difficulties

continue, contact RKI for further
instruction.

Advertising
This manual is related to the following products: