RKI Instruments 65-2649RK-HC-04 User Manual

Page 17

Advertising
background image

65-2649RK-HC-04 M2A Transmitter Operator’s Manual

12

12. Pull out the detector terminal strip and connect the four wires to the terminal strip as

follows (see Figure 6).

Connect the wire corresponding to the detector’s red wire to the LEL “R”
terminal.

Connect the wire corresponding to the detector’s white wire to the LEL “W”
terminal.

Connect the wire corresponding to the detector’s green wire to the LEL “G”
terminal.

Connect the wire corresponding to the detector’s black wire to the LEL “B”
terminal.

CAUTION:

If shielded cable is used, leave the cable shield’s drain wire disconnected and
insulated at the detector junction box. You will connect the opposite end of the cable’s
drain wire to the M2A junction box’s chassis (earth) ground.

Figure 6: Wiring the Detector to the M2A

13. Reinstall the detector terminal strip into its socket.

Red

W hite

Green

Black

Cable

Shield

Detector

Terminal

Strip

Green
White

Black

Red

Detector J-Box

M2A
J-Box

IR LEL Detector

Cal ibration Adapter/Splash G uard

C

NC

NO

AL

AR

M

1

C

NC

NO

AL

A

R

M

2

C

NC

N

O

FA

IL

-

S

+

A

B

C

RS

4

8

5

PW

R

/S

IG

R

W

G

B

LE

L

Advertising