RKI Instruments 65-2649RK-HC-04 User Manual

Page 35

Advertising
background image

65-2649RK-HC-04 M2A Transmitter Operator’s Manual

30

Flickering
Display

The display reading
flickers often.

• The M2A is

experiencing false

readings due to RFI

or EMI.

• The noise filter

setting is too low.

• The zero

suppression setting

is too low.

• The display screen

is malfunctioning.

1. Verify that the M2A wiring is

properly shielded. See “Wiring

the M2A Transmitter” on page 11.

2. Verify that the detector wiring is

correct and secure.

3. Increase the filter setting in

Configuration Mode.

4. Increase the zero suppression

setting in Configuration Mode.

5. If the display difficulties continue,

contact RKI for further

instruction.

Fail
Condition

• M2A indicates a

fail condition.

• Controller

indicates a fail

condition

• The detector wiring

to the terminal PCB

is disconnected or

misconnected.

• The wiring from the

M2A to the

controller is

disconnected or

misconnected.

• The detector fresh

air signal is low

enough to cause a

fail condition.

• The detector is

malfunctioning.

• The terminal or

display PCB is

malfunctioning.

1. Verify that the detector wiring is

correct and secure.

2. Verify that the wiring between the

M2A and the controller is correct

and secure.

3. Calibrate the M2A.
4. If the fail condition continues,

replace the detector.

5. If the fail condition continues,

contact RKI for further

instruction.

Slow or No
Response/
Difficult or
Unable to
Calibrate

• M2A responds

slowly or does not

respond to

response test.

• Unable to

successfully set

the fresh air or

span reading

during

calibration.

• M2A requires

frequent

calibration.

Note:

See “Calibration

Frequency” on page 31 for
calibration frequency
guidelines.

• The calibration

cylinder is low, out-

dated, or defective.

• The calibration gas

flow rate is too low.

• The calibration gas

is not an

appropriate

concentration.

• The detector is

malfunctioning.

• The terminal PCB is

malfunctioning.

• The flexible tubing

to the calibration

adapter/splash

guard has become

kinked, blocked, or

disconnected.

1. Verify that the calibration cylinder

contains an adequate supply of a

fresh test sample.

2. Verify that the regulator used for

calibration is a 0.5 LPM regulator.

3. Verify that the calibration gas

concentration is appropriate for

the transmitter. The concentration

should be in the detection range,

preferably about half of the

detection range.

4. Verify that the flexible tubing to

the calibration adapter/splash

guard is clear and connected to

the calibration adapter/splash

guard.

5. If the calibration/response

difficulties continue, replace the

detector.

6. If the calibration/response

difficulties continue, contact RKI

for further instruction.

Table 7:Troubleshooting the IR LEL Detector (Continued)

Condition

Symptom(s)

Probable Causes

Recommended Action

Advertising