4 How to Use

4-4. How to wire

- 19 -

TAN-5000

CAUTION

The specified bare wire length must be observed when the wire insulation is peeled off.
Improper clamping of the wire due to a shorter bare wire length may cause defective electric
conduction or heating.
Catching the wire insulation due to a shorter bare wire length may cause defective electric
conduction or heating.
Exposing the wire due to a longer bare wire length may cause defective insulation or a short circuit.
Be careful not to break up the wire.If the wire is broken up when inserted to the terminal, this may
cause defective insulation or heating.

Connection conditions
• Cable:

0.08 - 2.5mm

2

• Bare wire length: 8 - 9 mm
• Connecting tools: Dedicated screwdrivers manufactured by WAGO and equivalent

(edge width 3.5 mm x 0.5 mm or less)












Compatible bar terminal
For a bar terminal, the following items are available.
• Bar terminal (ferrule): Model 216 Series (manufactured by WAGO)

• Crimping tool: Model VarioCrimp 4 (206-204) (manufactured by WAGO)







• Dedicated

products

210-120J:.....Standard model
210-350/01:..Short model
210-258J:.....Angle model

• When using a general-purpose

screwdriver, use one with an edge width of
from 2.5 mm to 3.5 mm. Do not use a
screwdriver that does not fit into the
screwdriver slot or cannot open the spring
properly.

8 - 9 mm

CAUTION

A bar terminal of the specified model must be used. Using other bar terminals invalidates the
warranty of the performance.