IV. MAINTAINING YOUR HYBEX
®
SYSTEM
F. Checking and Replacing Fuses
Turn the power switch to the OFF position and unplug the
power cord before performing any service procedure.
There are two fuses located in a removable fuse block below the power cord receptacle on
the back of the unit.
Back of unit
Use small screwdriver behind
tab to open Fuse Block
Fuse Block
Push clip up to
release Fuse Holder
Digital Thermometer Jack
To remove fuses:
1. Insert a small, flat blade screwdriver into the tab recess just below the plug
receptacle.
2. Push down to release the fuse block.
3. Slide out the fuse holder from the fuse block while holding the retaining tab out of
the way.
4. Gently pry out the fuses.
A blown fuse appears dark. Always replace fuses with those of the same amperage and
voltage as shown on the label below the fuse block.