Hybex® Microarray Incubation System — USER MANUAL

Version 1.1E, August 2012 11 (408) 733-7337, [email protected]

V. TROUBLESHOOTING

Symptom

Cause

Solution

Power switch light does not turn on.

Blown fuse(s)

Replace fuse(s) on back of unit,
beneath power cord receptacle.

VI. SPECIFICATIONS

Electrical

Cat. #1057-36-1

115V AC; 50/60 Hz; 3A

Cat. #1057-36-2

230V AC; 50/60 Hz; 1.6A

Dimensions

Outside (cover closed) H x W x D

10x6x12 inches (21x15x30 cm)

Weight

Instrument (without block)

6 lbs (3 kg)

Performance and Controls

Temperature Range

Ambient +5°C to 99°C

Temperature Regulation

± 0.2°C

Temperature Controller

Digital PID, single loop

Temperature Display

Actual or Set single LED

Digital Thermometer Output

Thermocouple

VII. ORDERING INFORMATION

Cat. #

Description

UoM

1057-36-1

Hybex® Microarray Incubation System, 115V.

Each

1057-36-2

Hybex® Microarray Incubation System, 230V.

Each

1057-30-0

Hybex® Microsample Incubator, heating base, 115V.

Each

1057-30-2

Hybex® Microsample Incubator, heating base, 230V.

Each

1057-38-0

Microarray Chamber Set. Includes 2x chambers, tools and accessories. Each

1057-31-0

0.2 ml tube block. Holds 96x0.2ml PCR tubes or 96 well PCR plate.

Each

1057-33-0

0.6mL tube heat block, holds 60x0.6mL tubes.

Each

1057-34-0

1.5 ml tube block. Holds 32x1.5ml microcentrifuge tubes.

Each

1057-39-0

Block for flat-bottom microplate.

Each

RP1057-32-1 Absorbent Pads for Chamber Cover, 25/pack.

Pack

1057-32-3

Hybex® Microarray Rack, holds 4 slides.

Each

1057-32-5

Handle for Duplexed Hybex® racks.

Each

1057-35-0

Waterbath Insert for Hybex® Incubator.

Each

1051-52-0

Digital thermometer. Includes cable and NIST certificate.

Each

Advertising