Changing the instrument setup
1. Open the Setup/Calibration software.
2. Enter your initials using the ‘Enter Initials’ command from the

‘Device’ menu (shortcut is F7).

. Press the ‘Get from instrument’ button at the bottom of the screen

(shortcut is F9).

. The screen will now show the instruments current configuration.
5. Save the current configuration to disk using the ‘Save as’ command

from the ‘File’ menu.

6. Setup the instrument to suit your application via the

input screen (press F5) and output screen (press F6).

7. Return to the main screen, check the details are correct and press

the ‘Send to Instrument’ button (shortcut F8).

Note: You are required to enter a password to change the instruments
setup. The default password is 100. Make a note of the instrument
password if you change it - otherwise you will have to send the
instrument back to us.
8. Save the changes to disk as a record of changes you have made.

Main screen

Fields

Description

ID Tag

User defined field (e.g., PT1015-2)

User reference

User defined field (e.g., Pump 1015 bearing)