2 input calibration points – Weidmuller ITX+ - Universal signal isolator / amplifier User Manual
Page 24

2
Input Calibration points
Range
Values
Notes
Volts/Thermocouple
10 mV, 20 mV,
50 mV, 80 mV,
200 mV, 1 V
If the internal switches are set to 10 V, you must supply
a signal ten times the displayed value, e.g. 10 V for the
1V point.
mA
20 mA
RTD/Resistance
100 Ω, 200 Ω,
500 Ω, 800 Ω,
2 kΩ, 10 kΩ
Use four wire mode
Individual calibration
As selected
Allows you to recalibrate any of the points above
Fine adjust
Current input
Fine tune the adjustment for the range you have selected.
All
All ranges
Complete calibration of all points. Internal switches must
be in the 1V position.
Input calibration
1. Save the current calibration to disk (select ‘Save Calibration data’
from the ‘Device’ menu).
2. Select ‘Calibrate Input’ from the ‘Device’ menu.
. Enter the calibration password (Default is 101).
. Choose the calibration required (see Input calibration points table for
guidance).
5. Supply the signals requested by the software and follow the
instructions on screen. Press the ‘Done’ button when finished.