Operation method, Rs485 communication function – Yamato Scientific BBL300 Low Constant Temperature Water Bath User Manual

Page 36

Advertising
background image

33

Operation Method

RS485 Communication Function

5. Communication Examples

5.1 Read communication example

Example) Request message:

A request for reading PV is transmitted to the unit set at address 02.
Reply message from the unit to this request message: The data of PV (00123) is returned.


S

T

X

0 2

R

P

V

1

E

T

X

B

C
C






Code Symbol/Data

ASCII

code

*2

Start Code

STX 02H

Address

02

30H 32H

Request (Read)

R 52H

Identifier

*1

PV1

50H 56H 31H

Numeric Data

00123

30H 30H 31H 32H 33H

End Code

ETX 03H

Request

61H

BCC data

Reply

02H

Acknowledgement code

ACK 06H

*1): See "7. List of Identifiers/Commands."
*2): For ASCII codes, see "8. List of ASCII Codes."

S

T

X

0

2

A

C

K

P

V

1

0

0

1

2

3

E

T

X

B

C
C

Read request message

(transmitted from the host computer)

Reply message (returned from the unit)