Operation method, Rs485 communication function – Yamato Scientific BBL300 Low Constant Temperature Water Bath User Manual

Page 40

Advertising
background image

37

Operation Method

RS485 Communication Function

Other Parameters

Name

Identifier

Command

Setting Value

Key lock

LOC

R/W

00000:Key lock released
00001:Key lock

Operation
start/stop

RUN R/W

00000:Stop (*3)
00001:Start

Operation type
selection

RST R/W

00000:Fixed temperature operation selected (*3)

Remaining hour
monitor

_TI R

00000:Time-up (*1)
00001∼09950:0 hours and a minute to 999 hours and 50
minutes

Output monitor

OM1

R

00000:First digit = Heater output

Second digit = Refrigerator output
Third digit = Main output
Fourth digit = Time-up or alarm output
Fifth digit = Overheat prevention output

Output state: 0 = Output OFF, 1 = Output ON

Error monitor 1

ER1

R

00000:First digit = Memory error

Second digit = Sensor error
Third digit = AT error
Fourth digit = Heater wire disconnection error
Fifth digit = SSR short error

Error state: 0 = No error exists., 1 = An error exists.

Error monitor 2

ER2

R

00000:First digit = Boil-dry error

Second digit = Overheating prevention 1 error
Third digit = Overheating prevention 2 error
Fourth digit

= Internal communication/Temperature input circuit
error

Fifth digit = Unused

*Error state: 0 = No error exists., 1 = An error exists.

Measured
temperature
monitor

PV1

R

(Example)
00100

= 100

°C (when the temperature sensor is a thermocouple

input)

01000

= 100.0

°C (when the temperature sensor is a platinum input)

HHHHH

= Measured temperature over-scale (input common)

LLLLL

= Measured temperature under-scale (input common)

*The measured temperature resolution of the platinum input is

ten times that of the thermocouple input.