Before Using This Unit

Requirements for Installation

WARNING!

1. Choose a proper place for installation

• Do not install this unit in a place where:

♦ Rough or dirty surface.

♦ Flammable gas or corrosive gas is generated.

♦ Ambient temperature above 30°C.
♦ Ambient temperature fluctuates violently.

♦ There is direct sunlight.

♦ There is excessive humidity and dust.

♦ There is a constant vibration.
♦ Winds from the air conditioner, etc. hit the sample container directly.

• Install this unit on a stable place with the space as shown below.

More than

20cm

More than

20cm

More than 20cm

More than 20cm

Main Unit

2. Installation on horizontal surface

• Place this unit as flat a place as possible. If the

four feet are not in uniform contact with the floor
surface, noise or vibration may result.
Additionally, the unit may cause a problem or
malfunction.

• Weight of unit is BBL100 type: 35kg BBL300

type: 55kg. Conveyance and installation should
be done carefully by two or more persons.

3. Before/after installing

• It may cause injure to a person if this unit falls down or moves by the earthquake and the

impact. etc..To prevent, take measures that the unit cannot fall down, and not install to busy
place.