Installation/Assembly

Glassware

9. Set the cooling condenser nut, coil ring and vacuum seal to the condenser

or condenser connector.

1.Remove the cooling condenser nut on the left

side of the motor.

2.Connect the cooling condenser nut and then

coil ring to the condenser or condenser

connector (which ever applies).

3.Insert the vacuum seal to the condenser or

condenser connector (which ever applies).

*It is optional to put grease on the vacuum seal.

10. Connect cooling condenser or condenser connector to the motor.

Insert the steam duct into vacuum seal, put glass flange to the motor and fasten firmly the cooling

condenser nut.

Be careful not to damage the vacuum seal when you insert steam duct into the vacuum

seal in case the damage could cause leak.

Glass Type A

Glass Type B/C

Above :

Fitting of Coil Ring and Vacuum Seal

Below :

Fitting to Motor

Above :

Fitting of Coil Ring and Vacuum Seal

Below :

Fitting to Motor

When you remove the coil ring from cooling condenser or condenser connector

Hook the coil ring by the attached hexagon wrench as shown in the left

picture in order to remove the ring easily from the cooling condenser or

condenser connector. However, be careful not to force too hard in case

the glass apparatus or coil ring might be damaged.

Coil Ring

Vacuum Seal

Coil Ring

Vacuum Seal