Field Calibration
Data can be checked for calibration and integrity by carrying out a test on a 
known mass, for example, a calibration weight or steel bar of a measured value.
Use a calibration weight of 1kg or any other known mass.
Figure 2 - Field Calibration Setup
This is an example of a set-up for Field Calibration.
1.
Fit the correct tip to the modally tuned hammer.
Indicates how many hits 
were taken.
FRF Measurement
The FRF is colored red 
where coherence falls 
below 80%.
Good coherence shows the 
trace in green. 
Phase
Coherence for the 
measurement, represented 
by the color of the FRF 
trace.