DC Resistance Measurement

Preparation

and Tools

Required

Be sure that the vent plug is installed.

Preparation:

s

Tools: Common hand tools

s

Ohmmeter with an accuracy of ± 1% at 50

ohms.

s

Temperature of the load between 20°C to 25°C

(68°F to 77°F)

VSWR and RF Impedance are the true benchmark of a

quality dummy load. Accurate measurement of the dc re-

sistance between the inner and outer conductors of the

RF input connector will provide a good check of the con-

dition of the load resistor. Checking the dc resistance is

simply used to measure a change in the condition of the

resistor over time. The tracking of the dc resistance must

start before the resistor is first put into service. Perform

the following steps and record the value for future com-

parison. Check and record the resistance of the load peri-

odically according to use.

WARNING

Never attempt to connect or disconnect RF equipment

from the transmission line while RF power is being

applied. Leaking RF energy is a potential health

hazard.

Checking the

DC

Resistance

Measure the dc resistance. Follow the instructions below.

1. Turn off RF power and interlock circuitry before

any electrical disconnection’s are made.

2. Disconnect RF coaxial line.

3. Connect the multimeter test leads across the cen-

ter and outer conductor of the load resistor. Refer

to figure 3.

4. Record the value of the resistance before the load

is put into service. Compare subsequent values

with the latest reading. If the values vary more

than two ohms this could be an indication of a

failing resistive element.