Please carefully review these instructions to ensure proper and accurate use of your new Digitronic Level.
Calibrating this Moore & Wright Digitronic Level.

...LEVEL

ZERO

HOLD

...LEVEL

ZERO

HOLD

...LEVEL

ZERO

HOLD

...LEVEL

ZERO

HOLD

...LEVEL

ZERO

HOLD

...LEVEL

ZERO

HOLD

...LEVEL

ZERO

HOLD

...LEVEL

ZERO

HOLD

...LEVEL

ZERO

HOLD

3. Release the buttons

and make sure the
level does not move
until the ‘1’ flashes.

5. Turn the Level

another 90 degrees
clockwise; press the
ZERO button, the
screen will read
CAD3, make sure
the level does not
move until the ‘3’
flashes.

7. Turn the box

another 90 degrees
clockwise; press the
ZERO button, the
screen will read
CAD5. After 3
seconds the display
will read 0.00, this
completes the
calibration.

2. Press and hold ZERO

and button for
5 seconds until the
screen displays CAD1.

4. Turn the level 90

degrees clockwise,
press the ZERO
button, the screen
will read CAD2,
make sure the level
does not move until
the ‘2’ flashes.

6. Turn the Level

another 90 degrees
clockwise; press the
ZERO button,the
screen will read
CAD4, make sure
the level does not
move until
the ‘4’ flashes.

8. Your Digitronic Level

is now ready to use.

9. Low battery warning.

1. Place the level on

a flat surface.
The accuracy of
the granite surface
must be

0.02º in

order to make the
recalibration
effective.

The Level should be set right side up.

The Level should be set right side up.

The Level should be set on its right side.

The Level should be set on its top.

The Level should be set on its left side.