2 mounting the dosing tubing and buret tips, Mounting the dosing tubing and buret tips, Figure 13 – Metrohm 807 Dosing Unit User Manual

Page 25: Mounting the tubing and the buret tip

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

807 Dosing Unit

■■■■■■■■

17

3.5.2

Mounting the dosing tubing and buret tips

6.1805.100

6.1543.060/
6.1543.200

6.2739.000

Figure 13

Mounting the tubing and the buret tip

Mount the tubing and the buret tip as follows:

1 Screw the dosing tubing 6.1805.100 (40 cm) on the left-hand side to

Port 1 of the dosing unit.

2 Tighten the connection nipple with the wrench 6.2739.000.

3 Screw the open dosing tip 6.1543.060 or the antidiffusion tip

6.1543.200 onto the dosing tubing.

The link stopper 6.1446.030 included in the scope of delivery can be
used to fix the tip in place in a ground joint opening NS14/15.

4 Insert the buret tip into the storage vessel provided for this purpose.

Details concerning the selection of the buret tip (see Chapter 3.5.3, page
18)
.

The first-time filling of the dosing unit requires no special measures. Each
Metrohm control device (e.g. Titrando, Titrino or Sample Processor) is
equipped with a PREP/Preparing function with which the filling of the
dosing cylinder and the rinsing of the dosing unit tubing can be readily
accomplished in an automated sequence.