3.7 Disassembling the dosing unit

■■■■■■■■■■■■■■■■■■■■■■

22

■■■■■■■■

807 Dosing Unit

Removing the housing

Figure 15

Removing the housing

Remove the housing of the dosing unit as follows:

1 Place the dosing unit on a flat, level surface.

The lettering of the volume specification faces towards the front.

2 Keep the locking button of the spring clip pressed down.

3 Rotate the housing by approx. 1 cm to the right (in counterclockwise

direction).

4 Release the locking button.

5 Carefully raise the housing upward.

You will now see the complete cylinder unit with the transparent
centering tube which rotates on the distributor with the interior dos-
ing cylinder when the flat stopcock is switched.