6.2

Dosing accuracy

Every dosing unit is subjected to a strict quality inspection prior to ship-
ment. Every dosing unit is issued a quality certificate attesting conform-
ance with the quality criteria of Metrohm.

6.2.1

Typical measurement deviation

The accuracy of dosing units can be seen in the following table. The val-
ues listed are to be regarded as typical values which can be achieved with
a 700 Dosino or an 800 Dosino.

Table 3

Typical measurement deviation of Metrohm dosing units

Cylinder volume

max. systematic deviation

2 mL

± 6 µL

5 mL

± 15 µL

10 mL

± 20 µL

20 mL

± 30 µL

50 mL

± 50 µL

6.2.2

The ISO/EN/DIN standard 8655-3

The Metrohm dosing units fulfil the requirements of the ISO/EN/DIN
standard 8655-3 Volume measurement instruments with pistons –
Part 3: Piston burets
. Metrohm guarantees that its dosing units are in
compliance with the following limit values at the time of shipment: