Metrohm Inline Dialysis User Manual

Page 17

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

IC equipment: Dialysis

■■■■■■■■

11

The acceptor solution is pumped with two pump tubings with yellow/
orange stoppers (6.1826.320). Additional information on the tubing con-
figuration of the peristaltic pump can be found in the chapter "Installing
the peristaltic pump" in the manual for your ion chromatograph
.

1 Preparing two pump tubings for the sample

Use the two pump tubings with black stoppers (6.1826.340) for con-
veying the sample. Proceed as follows for each tubing:

■

Attach the coupling olive/UNF 10/32 (6.2744.034) to the inlet.

■

Attach the pump tubing connection with locking nut
(6.2744.160) to the outlet (see chapter "Installing the peristaltic
pump" in the manual for the ion chromatograph or in the man-
ual for the Sample Processor)
.

2 Preparing two pump tubings for the acceptor solution

Use the two pump tubings with orange/yellow stoppers (6.1826.320)
for conveying the acceptor solution. Proceed as follows for each tub-
ing:

■

Attach the coupling olive/UNF 10/32 (6.2744.034) to the inlet.

■

Attach the pump tubing connection with locking nut
(6.2744.160) to the outlet (see chapter "Installing the peristaltic
pump" in the manual for the ion chromatograph or in the man-
ual for the Sample Processor)
.

3 Connecting the sample inlet

■

Tighten the PEEK capillary, 0.5 mm ID / 70 cm (6.1831.160), to
the inlet of the pump tubing with black stoppers (6.1826.340)
using a pressure screw (6.2744.070).
Connect the other end of the PEEK capillary to the needle of the
Sample Processor (see manual for the Sample Processor).