4.2 Maintenance

■■■■■■■■■■■■■■■■■■■■■■

20

■■■■■■■■

IC equipment: Dialysis

4.2

Maintenance

The dialysis membrane may need to be replaced:

■

if the signal intensity during dialysis decreases.

■

if the dialysis membrane dries out.

■

if the dialysis membrane becomes damaged as a result of depositions
or bacterial growth.

■

if the sample channel is blocked (i.e., the sample can no longer be
pumped through the dialysis cell).

Replacing the dialysis membrane

1 Disassembling the dialysis cell

■

Unscrew the four capillaries from the dialysis cell and remove the
cell from the cell holder.

■

Remove the five screws completely with the hex key (6.2621.070)
and disassemble the dialysis cell.

■

Remove the used dialysis membrane.

2 Preparing the dialysis membrane

Carry out steps 2 to 8 in Chapter 3.1 Preparing the dialysis cell on
page 6
.

3 Connecting the dialysis cell

■

Tighten the sample inlet capillary to the dialysis cell inlet for the
sample (3-4) using a PVDF pressure screw (6.2744.000).

■

Tighten the sample outlet capillary to the dialysis cell outlet for
the sample (3-3) using a PVDF pressure screw (6.2744.000).

■

Tighten the acceptor inlet capillary to the dialysis cell inlet for the
acceptor solution (3-5) using a PVDF pressure screw (6.2744.000).

■

Tighten the acceptor outlet capillary to the dialysis cell outlet for
the acceptor solution (3-6) using a PVDF pressure screw
(6.2744.000).

4 Conditioning the dialysis system

(see Chapter 3.4, page 15).