5 Storing the electrode

The type of storage depends on the storage duration. Observe the follow-
ing notes:

Storage between measurements

Store electrode in distilled water.

Storage duration between 1 and 5 days

Store the electrode in measuring electrolyte (6.2316.030).

Storage duration of more than 5 days

1 Unscrew membrane module.

2 Rinse the membrane module thoroughly with distilled water inside

and outside and place it in dry storage.

3 Store pH glass electrode in c(KCl) = sat. (6.2308.000).