3 performing direct measurements, Performing direct measurements – Metrohm NH3-selective electrodes User Manual

Page 16

Advertising
background image

6.3 Performing direct measurements

■■■■■■■■■■■■■■■■■■■■■■

10

■■■■■■■■

NH

3

-selective electrodes

■

The electrode must be conditioned in distilled water for at least 10
minutes between the measurements. If measurements are being per-
formed at very high concentrations, then the electrode should be con-
ditioned for 30 minutes.

■

If a standard or sample solution is measured repeatedly, then it must
be stored in a sealed or covered vessel between the measurements.
This ensures that ammonia will not escape.

■

Ammonia dissipates very rapidly out of solution when ammonia con-
centrations are > 1 mol/L. These kinds of samples must therefore be
diluted.

■

You will find many additional useful notes on working with the NH

3

-

selective electrode in Application Bulletin 133.

6.3

Performing direct measurements

Direct measurement is a simple and rapid method for testing numerous
samples across a wide range of concentrations.

Observe the following notes when performing direct measurements:

■

Select the concentrations of the standard solutions (e.g. NH

4

Cl solu-

tions) in such a way that the ammonia concentration of the sample
solution to be anticipated falls in the middle of the calibration range.

■

Make sure that the temperatures of all standard and sample solutions
are identical.

■

The diffusion of ammonia through the membrane is slowed down con-
siderably in the presence of ammonia concentrations < 6 · 10

–5

mol/L.

The response time of the electrode is prolonged accordingly. A special
measurement technique is required for this concentration range (see
Chapter 6.5, page 12)
.

■

Ammonia dissipates very rapidly out of solution when ammonia con-
centrations are > 1 mol/L. These kinds of samples must therefore be
diluted (see Chapter 6.6, page 13).

Advertising