6 performing measurements, 1 measured quantities, 2 general notes – Metrohm NH3-selective electrodes User Manual
Page 15: Measured quantities, General notes

■■■■■■■■■■■■■■■■■■■■■■
6 Performing measurements
NH
3
-selective electrodes
■■■■■■■■
9
6 Performing measurements
6.1
Measured quantities
Ammonia concentrations are usually specified in one of the following
units:
■
mol/L
■
ppm (NH
3
)
■
ppm (N)
Table 2
Conversion factors for various concentration specifications
Conc. in mol/L
N content in ppm
NH
3
content in ppm
10
–4
1.4
1.7
10
–3
14
17
10
–2
140
170
10
–1
1400
1700
1
14000
17000
6.2
General notes
■
Standard and sample solutions must always be measured at the same
stirring speed and at the same temperature. A difference in tempera-
ture of 1 °C results in a deviation of approx. 2%.
■
If possible, use narrow, high measuring vessels (minimum ratio of sur-
face to volume).
■
Immediately prior to the measurement, a sodium hydroxide solution
must be added to each measuring solution. The solutions should have
a pH value of 11...14 after this addition and the concentration of all
dissolved particles should not exceed 1 mol/L.
■
Alkali samples must be measured immediately. The ammonia loss in a
stirred alkali sample of 100 mL is approx. 50% within six hours.
For purposes of storage, the samples must be acidulated with hydro-
chloric acid to approx. pH 6 (approx. 0.5 mL c(HCl) = 1 mol/L per liter
of sample) and stored in well-sealed vessels.
Do not add the sodium hydroxide solution until just before the mea-
surement.