11

3. Set the detector temperature to 100°C and allow time for the detector and

helium purifier to reach the set temperature.

4. Set the range to 10

2

and adjust the voltage to Ø.

5. Plug the 15 VDC output from the power supply (I-23569-1) into the pulser

module (PD-M2).

6. Plug the power supply cord to a 100 - 250 VAC source, and look through

the hole in the detector housing to see that the discharge is on.

CAUTION: During normal operation, the detector pro-
duces ultraviolet energy (UVA, UVB), some of which may
be emitted. Do not watch the arc without eye protection.

7. Check the standing/background current. Optimum range is 6.0 - 20.0 mV

(0.60 - 2.0 nA) at 100°C in the HID mode. Lower current indicates a
clean, leak-free system. Record the standing/background current in the
Detector Performance Log on the last page of this manual.

8. The recommended detector temperature is 20°C above the column

temperature, with a minimum of 100°C. Set the detector to the operating
temperature required for the intended analysis. When the detector has
reached the set temperature, read and record the standing current.

9. Install the column as illustrated on page 12, leaving the oven at ambient

temperature. Start carrier flow, then read and record the standing current.

The difference between this reading and the one previous is the ionization
of the combined impurities in and eluting with the carrier gas. The smaller
the difference, the better the quality of the gas exiting the column.

10. Set the column oven to the temperature required for the intended

analysis. When the oven reaches the set temperature, read and record
the standing current.

The difference with the previous reading is the ionization of the column
bleed. The smaller the difference, the better the column is conditioned.

NOTE: Some stationary phases will have a higher bleed than others,
but are still suitable for this detector. However, the lower the bleed, the
lower the chances of contaminating the detector cell.

From this point, the standing current should be observed and logged after
any system change. In addition, logging the standing current (with and
without the column) on a regular basis is an effective monitor of system
integrity (leaktightness and cleanliness). We also recommend tracking the
internal standard (quantity on column/area count) for sensitivity continuity.