Reset av38, Performance specification – LaMotte 2020 ClearTrace User Manual

Page 32

Advertising
background image

12. RESET AV38

To reset all configuration
information to default
values, select “Reset AV38”
from the main menu.

The AV38 will ask the
question, “Are you sure?” to
notify the user that they are
about to make changes to
the system. Press ENTER to
initiate a reset. Press ESC to
abort the reset.

When the reset is done, the
AV38 will require that the
Enter key is pressed to be
sure the user acknowledges
that a reset has been done.

13. PERFORMANCE SPECIFICATION

Measurement
Range: 0 to 100 NTU (Can range to 200 NTU when necessary).
Resolution: 0.001 NTU
Accuracy: +/-2% of reading or +/-0.015 NTU whichever is greater. +/-5%
of reading above 40 NTU.

Operational Environment
Water Temperature Range: -5°C to 50°C
Air Temperature Range: -20°C to 60°C
Maximum Flow Rate: 500 ml/min (7.9 gal/hr)
Minimum Flow Rate: 250 ml/min (4 gal/hr)

31

Reset AV38
Done

MENU

ESC

ENTER

t

t

t

t

Main Menu
Reset AV38

MENU

ESC

ENTER

t

t

t

t

t

Reset AV38
Are you sure?

MENU

ESC

ENTER

t

t

t

t

Advertising