LaMotte 2020 ClearTrace User Manual
Page 33
Power Requirements
Voltage Range: 24 VDC(Line power option available)
Maximum Power: 8W with AV38 and Light Source
Typical Power: 6W with AV38 and Light Source
Note: Class II DC power supply required
Construction
Light Source: White Light (Tungsten)
Sample Chamber Material: ABS plastic
Sample Chamber Volume: 135 ml
Light Source Housing Material: Anodized Aluminum
Mounting Plate: 12x12 inches, 4 mounting holes
Sensor Head Material: Quartz Glass, Anodized Aluminum
Weight: 5.6 lbs
Units of Measure
Measurement Units: NTU
Temperature Units: °C, °F
Calibration
Sample (Span): 1 point
Zero: 1 point
Temperature: 1 point
Interface
Display: 2 lines, 16 characters, 7 key menu navigation
Current Loops: 1 loop standard.
Interface
Sensor Filter: 0 to 100 Seconds
Temperature Filter: 0 to 100 Seconds
Product Safety
cULus Listed 367G E 328889
32