LaMotte pH PLUS Direct Meter User Manual

Page 11

Advertising
background image

3.2 Selection of pH Buffer Standards
You must set the meter to accept either USA, NIST (nSt) or Low Ionic (Pb)
pH buffer standard values before calibration. The factory default is USA
standard. If you wish to abort this operation press the

CAL key at any sequence

and the meter reverts to pH measurement mode.
1. Press and hold the

MODE key. Switch on the meter using ON key. The

display shows “

bUF” blinking.

2. Press the

ENTER key to get into the buffer selection mode. Use the

MODE key to toggle between USA, NIST or Pb standards as shown.

3. Press the

ENTER key to confirm the buffer standards to be used. The

display then reverts to pH measurement mode.

3.2.1 Resetting the User Calibrated Values

Note: only the temperature offset (if set) will not be erased.

1. Press and hold the

CAL key while switching on the meter using the ON

key. The LCD shows “

rSt” blinking.

2. Press the

MODE key to abort this operation if you do not wish to reset

3. Press the

ENTER key to confirm. The meter automatically clears all stored

pH/Ion calibration or mV offset values and reverts to measurement mode.

11