LaMotte pH PLUS Direct Meter User Manual

Page 18

Advertising
background image

5. To cancel or abort this operation, press the

CAL key. Note no new value

will be stored into the meter's non-volatile memory. To confirm the
calibration, press the

ENTER key. The LCD displays “CO” momentarily,

and the meter reverts to the measurement mode.

3.4.2 Without Temperature probe (no ATC)
If a temperature probe is not used, the meter compensates for the pH response
based on a new calibrated temperature value manually set or at 25.0 °C
(factory default).
1. Press the

MODE key to enter into Temperature mode until “°C” shows on

the LCD.

2. Compare the displayed value to a NIST certified thermometer or

thermometer known to be accurate (dipped into a constant temperature
bath).

3. Press the

CAL key to enter the temperature calibration mode. The LCD

shows “

CA” momentarily and the displayed reading flashes. Note that this

displayed value should either be 25.0

o

C or the last set temperature value.

4. Press

p and q keys until the display shows the desired temperature. Any

value from 0 to 100

o

C can be set.

5. To cancel or abort this operation, press the

CAL key. Note no new value

will be stored into the meter‘s non-volatile memory. To confirm the
calibration, press the

ENTER key. The LCD displays “CO” momentarily,

and the meter reverts to measurement mode.

18