LaMotte pH PLUS Direct Meter User Manual

Page 20

Advertising
background image

5. ELECTRODE CARE AND MAINTENANCE
For best results, always keep the ISE capped dry and pH/ORP electrode bulb
wet. Store the pH/ORP glass bulb with pH electrode storage solution. pH
buffers are also suitable.

NEVER use deionized water for storage. Wash the

probes thoroughly with distilled water after each use. Because the ISE or pH
electrode is susceptible to contamination or dirt, clean it every 1 to 2 months
depending on extent and condition of use.
Clean the pH/ORP electrode using a mild detergent. Wipe the probe with soft
tissue paper. Avoid touching the glass membrane with your fingers. Wash the
electrode thoroughly in tap water and then in distilled water. Recalibrate the
meter after cleaning the electrode.

6. TROUBLESHOOTING

Problem

Cause

Solution

No display

Batteries not in place

a) Insert batteries.
b) Re-insert batteries with correct polarity.

LO” displays in

the LCD

Low battery

Replace batteries with fresh ones.

Unstable
reading

a) Electrode not deep
enough in sample
b) Dirty electrode
c) Broken electrode

a) Place electrode deeper in sample.
b) Clean electrode and recalibrate.
c) Replace electrode.

Er1” display

Buffer value out of
tolerance

Use new pH buffer solution and recalibrate.

Er2” display

Single point

Perform at least 2 point calibration.

Er3” display

ISE slope not within
the specified tolerance

Check ISE is in good working condition

.

Er4” display

Calibration points not
within 1 decade

Ensure calibration points are at least 1
decade apart.

Not able to
calibrate

a) Display freezes
b) Faulty electrode

a) Release reading by pressing HOLD key.
b) Replace electrode.

20

Advertising