QUANTITY

CONTENTS

CODE

30 mL

Ammonia Nitrogen Reagent #1

4797-G

30 mL

*Ammonia Nitrogen Reagent #2

*4798-G

120 mL

*Mixed Acid Reagent

*V-6278-J

10 g

*Color Developing Reagent

*V-6281-D

30 mL

*Wide Range Indicator

*2218-G

15 mL

*Salinity Indicator Reagent A

*7460-E

60 mL

*Salinity Titration Reagent B

*7461-H

30 mL

*Manganous Sulfate Solution

*4167-G

30 mL

*Alkaline Potassium Iodide Azide

*7166-G

50 g

*Sulfamic Acid Powder

*6286-H

60 mL

Sodium Thiosulfate, 0.025N

4169-H

30 mL

Starch Indicator Solution

4170WT-G

4

Square Tubes, plastic, w/cap

0106

1

Test Tube, glass, 5-10-15 mL, w/cap

0778

1

Test Tube, 5-10-12.9-15-20-25 mL,
w/cap

0608

1

Direct Reading Titrator, 0-10 Range

0377

1

Direct Reading Titrator, 0-20 Range

0378

1

Direct Reading Titrator, 0-1 Range

0376

1

Pipet, 0.5 mL, plastic

0353

1

Spoon, 0.1g, plastic

0699

1

Spoon, 1.0g, plastic

0697

1

Nitrite Nitrogen ColoRuler, 0.5-2.0 ppm 3551

1

Ammonia Nitrogen Color Chart,
0.5-2.0 ppm

3560-CC

1

pH Color Chart, 5.0-8.0 pH

3603-CC

1

Water Sampling Bottle, 60 mL, glass

0688-DO

1

Demineralizer Bottle

1151

1

Thermometer

1066

*WARNING:

Reagents marked with a * are considered to be potential health

hazards. To view or print a Material Safety Data Sheet (MSDS) for these reagents
see MSDS CD or www.lamotte.com. To obtain a printed copy, contact LaMotte
by email, phone or fax.

To order individual reagent refills or test kit components, use the specified code
numbers.

SA LIN ITY TEST

1.

Fill test tube (0778) to 10 mL line with deionized water.

2.

Fill the Direct Reading Titrator, 0-1 Range (0376) with sample water.
Wipe any excess sample water from the tip.

3.

Dispense 0.5 mL of the sample water into the test tube by slowly
pressing the plunger until the tip of the plunger is opposite the 0.5
mark. Discard the remaining water in the Titrator.

4.

Add 3 drops of *Salinity Indicator Reagent A (7460) to the test tube.
Cap and gently swirl to mix the contents.

5.

Fill the Direct Reading Titrator, 0-20 Range (0378) with *Salinity
Titration Reagent B (7461). Insert into the center hole of the test tube
cap.

6.

While gently swirling the test tube, slowly press the plunger to titrate
until the color of the solution changes from yellow to pink-brown.
Read the test result directly from the scale where the large ring on the
Titrator meets the Titrator barrel. Record as ppt salinity. Each minor
division equals 0.4.
EXAMPLE:

Plunger tip is 3 minor divisions below line 10. The test

result is 10 plus (3 divisions x 0.4) equals 11.2 ppt.