Test Kit Procedure 125 continued
NOTE 1:
Step 5 is very important to the accuracy of the titration and should always be
performed. The reason for this is due to the natural occurring alkalinity in the
water used to dilute the Oakite Product. This alkalinity will give erroneously
higher concentration values unless corrected.
*NOTE 2:
For complete instructions on how to use the Direct Reading Titrator please
refer to the enclosed instruction sheet.
Use Concentration
Sample Volume
Multiplying Factor (F)
Oakite Material
g/
l
(oz/gal)
% by vol
in ml
% by vol
g/
l
(oz/gal)
53
3.8-15(0.5-2)
2
8.2 (1.1)
101
2-8
5
4.9
202
1-4
10
4.2
FiSan-LC66
0.5-2
10
0.87
FiSan-LC99
0.5-1.5
10
0.93
FiSan-Panacea
1-4
5
1.4
Fleetline JC5
5-20
2
9.1
Fleetline 170
7.5-30 (1-4)
2
22 (2.9)
Fleetline 205
5-20
2
11
Fleetline 205
Special
5-20
2
11
FM 403
1-4
5
1.6
Liqui-Det
2-8
5
3.7
Liqui-Det 2
0.5-2
10
2.0
Liquid Power-Det
2-8
2
5.0
Solventol
Hydra-Clean
3-10
1
6.2
Versadet
3.8-15 (0.5-2)
1
11 (1.5)
7/26/83
Supersedes 11/20/80
PTD. in U.S.A.