Device Placement (Bottom View)
NOTE:
1.
Two electrical/communication/medical gas devices cannot be placed together at any corner.
2.
Maximum of 12 gas outlets are allowed on the column. If 12 gas outlets are required, then side mount is optional.
3.
All electrical devices must be located within section A, B, C, or D only.
4.
The Nitrogen Control Panel will occupy three spaces of the bulkhead (E.G. A2 to A4 see Outline).
Location of Nitrogen Control Panel
Side A
Side B
Side C
Side D
Please Specify Outlet Language
US
CAN
FRENCH
SPANISH/US
SPANISH/ISO
Gas Legend
Oxygen
= O
Medical Air
= A
Vacuum
= V
Nitrous Oxide = 2
Nitrogen
= N
Carbon Dioxide = C
WAGD (US)
= W
AGSS (Canada) = E
Nitrogen Control Panel
Yes
o
No
Please fill the devices that you require for your application.
(Specify by using the symbols on Page 5 of this Approval Booklet.)