Description of the alarm, Description of the modules – Amico Nema 4 Master Alarm User Manual

Page 7

Advertising
background image

www.amico.com

7

Description of the Alarm

SHIPMENT DETAILS

When you receive an ALERT-2 series alarm from Amico Corporation, the package will consist of two main sections; the

Alarm Back Box and the Frame/Module Assembly. The Frame/Module assembly will be pre-configured, with the appro-

priate display modules, based upon the customer’s specifications.

THE ALARM BACK BOX

The Alarm Back Box contains the auto-switchable System Power Supply with an ON/OFF switch, a built-in fuse and ter-

minal blocks (115 VAC, or 220 VAC). The back box also incorporates the pipe stubs for applications that require locally

(in box) mounted sensors.

THE FRAME/MODULE ASSEMBLY

The Frame/Module Assembly consists of the frame and all the modules that are pre-assembled to the customers speci-

fication. The hinged frame is designed to swing down from the back box to facilitate installation and servicing of the

alarm. This design will reduce installation time and eliminate the risk of improper installation since all the modules are

connected and tested at the factory.

The ALERT-2 alarm is divided into 4 or 7 main modules:

COMMON TO ALL ALARMS

1. SYSTEM POWER SUPPLY

The System Power Supply has been pre-installed into the back box assembly. The System Power Supply converts the

AC voltage supply to the alarm into two voltages: 5 VDC (regulated) required by the microprocessor hardware and 15

VDC (unregulated) required by the buzzer and the LED’s. This unit also contains the main ON/OFF power switch, the

transformer, the heat sink, the main fuse and fuse cover, the rectifying circuitry, the terminal blocks and the low voltage

DC power cable for connecting this unit to the annunciator module. The System Power Supply can be easily removed

and reinstalled by unscrewing it from the back box.