Bio-Rad S3™ Cell Sorter User Manual
Page 111
Advertising
Instruction
Manual | 101
10.4 Swap Nozzle Tip Wizard
26. If after reseating the nozzle tip the stream is still hitting the back wall, sort tray, or deflection
plates, proceed to step 28.
27. Click the Stream Adjustment icon (see bottom of Figure 119).
28. The Adjust Stream Position window will open (Figure 121).
Fig. 120. AutoGimbal Swap Tip window.
Fig. 121. Adjust Stream Position window.