Bio-Rad Helios® Gene Gun System User Manual

Page 18

Advertising
background image

Attaching the Helios Gene Gun to the Helium Regulator

Components needed

Helium regulator attached to a helium cylinder

Helium hose assembly

Helios Gene Gun

Procedure

1. Insert the stem of the Swagelok Quick-Connect fitting on the helium hose into the opening in

the body of the Swagelok Quick-Connect fitting on the helium regulator and push until it
clicks. The helium hose will be locked into the helium regulator (see Figure 10). Note: If the
helium regulator has been pressurized, the stem and body will not lock. Turning the regulator
valve counter-clockwise will depressurize the system.

2. In a similar manner, insert the stem of the Swagelok Quick-Connect fitting on the Helios

Gene Gun into the opening in the body of the Swagelok Quick-Connect fitting on the
helium hose until it clicks. The Gene Gun will be locked into the helium hose.

Fig. 10. Connecting the helium hose to the helium regulator.

4.3 Setting up the Tubing Prep Station

Refer to Section 3.2, Identification of System Components and Controls, prior to system

installation. See Figure 11 for a diagram of the assembled Tubing Prep Station.

The Tubing Prep Station is shipped disassembled. The following sections describe assem-

bly of the Tubing Prep Station, attachment of the syringes and tubing, installation of the nitrogen
pressure regulator, and connection of the Tubing Prep Station to the nitrogen regulator.

A peristaltic pump is recommended for removal of the ethanol from the Gold-Coat tubing

after the microcarriers have been loaded (see Section 5.1). If a peristaltic pump is not available,
this may be done manually using a syringe. Assembly of the syringe and tubing is described
in step 5 in the following section.

14