Problem Cause Solution

b. Ionic strength of sample

b. Use same buffer in

lower than that of gel

sample as in gel or
stacking gel.

8. Skewed or distorted bands

a. Poor polymerization

a. Degas stacking gel

around the sample wells

solution thoroughly
prior to casting;
increase ammonium
persulfate and TEMED
concentration by 25%,
wipe comb with
TEMED just before
casting the stacking gel,
insert comb carefully to
prevent trapping of air
bubbles in sample well.

b. Salts in sample.

b. Remove salts by

dialysis, desalting
column, etc.

c. Uneven gel interface

c. Decrease the

polymerization rate.
Overlay gels very
carefully.

9. Lanes constricted at the

a. Ionic strength of sample

a. Desalt sample and

bottom of the gel.

higher than that of

neighboring samples.

surrounding gel.