Problem
Possible Cause
Remedy
Bands or spots not
No protein on gel
Verify that there is actually
visible
protein on the gel by
staining with another
method (e.g. Bio-Safe™
Coomassie stain).
Malfunctioning imaging
Check instrument manual
system
for troubleshooting.
Poor staining sensitivity
Residual SDS in gel
Confirm the correct amount
of fix solution is used. (See
page 5)
High or uneven
Dirty equipment or
Make sure that the staining
background staining
staining trays
trays and other equipment
used have been thoroughly
cleaned with laboratory
glassware cleaner. Use
glass staining trays, as
these are more likely to be
thoroughly clean.
Reagent impurities
Make sure that the water
and reagents used for
preparing and fixing the gel
are of the highest possible
quality. Proteomics grade
water may be used to pre-
pare the stain solution.